| Catalog Number |
SSF-605 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
03/20/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side rupture.The device has been explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side rupture.The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: d9, h3, h6.Device evaluation: based on the product analysis performed, the assessments of the complaint codes are: ¿ rupture: observed an opening assessed as surgical damage.As per the investigation procedure, creases, was observed and none of the observations are found to be potentially related to the manufacturing process, no further actions are required.
|
| |
|
Search Alerts/Recalls
|