U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENDOLOGIX AFX; VELA SUPRARENAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOLOGIX AFX; VELA SUPRARENAL Back to Search Results
Model Number A28-28/C95-O20V
Medical Device Problem Code Lack of Effect (4065)
Health Effect - Clinical Codes Aneurysm (1708); Clinical Endoleak (4843)
Date of Event 04/01/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx with duraply.
 
Event or Problem Description
The patient was treated for an abdominal aortic aneurysm on (b)(6) 2017 with the implant of an afx2 bifurcated stent graft and an afx vela suprarenal.Routine follow-up performed on (b)(6) 2025 identified a type ia endoleak.Reintervention was completed on (b)(6) 2025 with the implant of an afx vela infrarenal.The type ia endoleak was sealed and the patient is reported to be doing well.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.Endologix could not perform an evaluation of the device as the device remains implanted in the patient.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the type ia endoleak and additional endovascular procedure complaints are confirmed.This is consistent with the reported adverse event/incident.The clinical assessment determined that there was evidence to suggest aneurysm enlargement of 20mm that was not included in the event as reported.This was discovered during review of the discharge summary dated 15 april 2025.Device, user, procedure or anatomy relatedness of complaint could not be determined.No procedure related harms were identified.The final patient status was reported as discharged home on postoperative day one in good condition.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Corrections: g3: awareness date - updated.H6: investigation finding codes - remove code 3233.H6: investigation conclusion codes - remove code 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX
Common Device Name
VELA SUPRARENAL
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer (Section G)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer Contact
gary kirchgater
2 musick
irvine, CA 92618
8009832284
MDR Report Key21860932
Report Number3011063223-2025-00040
Device Sequence Number11933333
Product Code MIH
UDI-Device Identifier00818009013637
UDI-Public(01)00818009013637(17)200727
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/27/2020
Device Model NumberA28-28/C95-O20V
Device Lot Number1724800-010
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/01/2025
Supplement Date Received by Manufacturer05/27/2025
Initial Report FDA Received Date04/17/2025
Supplement Report FDA Received Date05/30/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/22/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX2 BIFURCATED STENT GRAFT (LN 1722975011).
Outcome Attributed to Adverse Event Required Intervention;
Patient Age75 YR
Patient SexFemale
-
-