• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ENERGEN? CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ENERGEN? CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) Back to Search Results
Model Number N141
Medical Device Problem Codes High impedance (1291); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/07/2019
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a retrospective review of device data it was identified that during a shock delivery, this system recorded a code 1005 indicative of high, out-of-range shock impedance measurements greater than 145 ohms.No other information was provided.This device was explanted.No adverse patient effects were reported.
 
Event or Problem Description
It was reported that during a retrospective review of device data it was identified that during a shock delivery, this system recorded a code 1005 indicative of high, out-of-range shock impedance measurements greater than 145 ohms.No other information was provided.This device was explanted.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
Follow up report 1 (device evaluation): the complaint device was not returned to boston scientific; therefore, product analysis could not be performed.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.A risk review was completed and confirmed that the event of device displays error message was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.At this point, it can be concluded that unknown factors most probably contributed to the event.Conclusion code was updated to "cause not established" because the reason for the alleged observation(s) could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENERGEN? CRT-D
Common Device Name
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key21861055
Report Number2124215-2025-24635
Device Sequence Number7917369
Product Code LWP
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P010012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2019
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/04/2013
Device Model NumberN141
Device Catalogue NumberN141
Device Lot Number103128
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/19/2025
Supplement Date Received by Manufacturer05/29/2026
Initial Report FDA Received Date04/17/2025
Supplement Report FDA Received Date05/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/06/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age66 YR
Patient SexMale
-
-