The device was not returned to nuvasive, and no photographs were provided so the complaint could not be confirmed.Review of the reported event noted the device tip fractured during usage and suggests off angled excessive force was incurred.Review of lot code information provided and device history record noted the device was released to the field on feb 24, 2014 as part of a field loaner set.Review of all the information received and similarly reported events suggests the root cause of the reported device jaw fracture the fracture was likely the result of material use fatigue combining with excessive forces to the articulating jaw and considered field damage resulting from wear fatigue and excessive force.These devices are part of loaner sets that are subjected to handling and repeated sterilization cycles after distribution into the field.Wear and damage can accumulate over time, no additional investigation required.Labeling review: "residual risks to the patient associated with use of general surgical instruments are cleaning and sterilization failures, sterile barrier compromise, corrosion, and unretrieved instrument fragments which may cause or contribute to infections, fever, wound dehiscence, immunological response and immunological, systemic, or allergic reactions.In some instances, treatment of these events may require surgical intervention including revision surgery." "instrument malfunctions which potentially result in surgical delays (thereby causing additional exposure to anesthesia, blood loss, and infection), a change in surgical plan, failure to achieve optimal clinical result and/or, infrequently, cancellation of a procedure.Such malfunctions include instrument fracture, which could make necessary removal difficult or sometimes impossible, with possible consequences of late infection and migration; instrument fracture may also result in injury to the patient." "residual risks to surgical staff associated with use of general surgical systems are: instrument failures or malfunctions which prevent intended use, resulting in decreased operational efficiency and general inconvenience." "patient education: preoperative instructions to the patient are essential.The patient should be made aware of the potential risks of the surgery." "pre-operative warnings the method of use for the instruments are to be determined by the user¿s experience and training in surgical procedures." "the instruments should be treated as any precision instrument and should be carefully placed on trays, cleaned after each use, and stored, according to generally accepted hospital methods and practices.The instruments should be carefully examined prior to use for functionality, excessive wear, or damage.A damaged instrument should not be used as this may increase the risk of malfunction and potential patient injury." "care should be used during surgical procedures to prevent damage to the device(s) and injury to the patient." "intra-operative warnings the physician should take precautions against putting undue stress on the spinal area with instruments.Any surgical technique should be carefully followed.It is important that the surgeon exercise extreme caution when working in close proximity to vital organs, nerves, or vessels, and that the force applied to the instrumentation is not excessive, to prevent potential injury to the patient.Over-bending, notching, striking, and/or scratching of implants with any instrument should be avoided to reduce the risk of breakage." "the physical characteristics required for many instruments do not permit them to be manufactured from implantable materials.If any broken fragments of instruments remain in the body of a patient, they could cause allergic reactions or infections.If an instrument breaks in surgery and fragments go into the patient, these pieces should be removed prior to closure and should not be implanted." "method of use if there is any doubt or uncertainty concerning the proper use of instruments, please contact nuvasive customer service.Any available surgical techniques will be provided upon request.For optimal results, the same type of instruments used for implantation should be used for implant removal." "information to obtain a surgical technique manual or other applicable instructions for use or should any information regarding the products or their uses be required, please contact your local representative or nuvasive directly at +1-800-475-9131.You may also email: info@nuvasive.Com." 9004421-en w-2022-12.
|