| Catalog Number |
CAT02462 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
03/21/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the anchor broke during the procedure and had to wait 30 mins for replacement.One (1) faulty anchor was unable to be removed and left inside the patient.
|
| |
|
Event or Problem Description
|
|
It was reported that the anchor broke during the procedure and had to wait 30 mins for replacement.One (1) faulty anchor was unable to be removed and left inside the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
It was confirmed the device is not available for investigation in-house.This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: anchor broke probable root cause: application - excessive force impacting inserter - extremely hard bone, no pilot hole drilled - poor patient bone quality the reported failure mode will be monitored for future reoccurrence.
|
| |
|