Model Number T14MH25
Medical Device Problem Codes Fracture (1260); Overheating of Device (1437)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 09/27/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the health care professional complained that the stylus motor was heating up during the surgery and the bur broke subsequently.A second bur ( same lot number) was used to replaced the broken bur, the same incident happened.A third bur was used to replace the broken bur and the hcp managed to finish the surgery.Broken pieces of the tool were found in the patient which were later retrieved.On follow-up it was reported that there was no patient or staff impact, and there was a delay of 10 minutes (2 times troubleshooting as 2 burs broke during the surgery) during delay burrs were replaced.It was also reported that there was no prolonged exposure to anesthesia and no intervention was performed.
 
Additional Manufacturer Narrative
H3: product analysis: evaluation determined that tool received used/ fractured at tip.The likely cause was identified as tool was applicated with too much force/ pressure causing it to fracture off; not sure what caused it to "over heat".This regulatory report is being submitted due to a retrospective review through capa 624392.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDAS REX LEGEND
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer (Section G)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key21863320
Report Number1625507-2025-01461
Device Sequence Number4592085
Product Code HBE
UDI-Device Identifier00763000061968
UDI-Public00763000061968
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K020069
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberT14MH25
Device Catalogue NumberT14MH25
Device Lot NumberH5836036
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2023
Initial Report FDA Received Date04/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/05/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-