| Catalog Number |
51-107120 |
| Medical Device Problem Code |
Manufacturing, Packaging or Shipping Problem (2975)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: 51-107170 item name tprlc 133 mp type1 pps ho 17.0 lot # 7244465 51-106150 item name tprlc 133 mp type1 pps so 15.0 lot # 6619595 51-105160 item name tprlc xr t1 pps 16x152mm lot # 7206037 51-103170 item name tprlc 133 t1 pps so 17x154mm lot # 7109654 51-106160 item name tprlc 133 mp type1 pps so 16.0 lot # 7004933 51-107150 item name tprlc 133 mp type1 pps ho 15.0 lot # 6121849 51-106160 item name tprlc 133 mp type1 pps so 16.0 lot # 6985721 51-107150 item name tprlc 133 mp type1 pps ho 15.0 lot # 6912866 51-106170 item name tprlc 133 mp type1 pps so 17.0 lot # 6639627 51-106170 item name tprlc 133 mp type1 pps so 17.0 lot # 7128190 51-107160 item name tprlc 133 mp type1 pps ho 16.0 lot # 6912870 51-107170 item name tprlc 133 mp type1 pps ho 17.0 lot # 6572017 51-100050 item name tprlc 133 fp type1 pps so 5.0 lot # 7391195 51-103130 item name tprlc 133 t1 pps so 13x146mmtp lot # 6644264 51-103150 item name tprlc 133 t1 pps so 15x150mm lot # 6974513 g2: foreign: japan customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that while inspecting the product, the sterile packaging was found damaged.There was no patient involvement.There is no further information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Product was returned and evaluated.Visual evaluation of the returned product/provided photos identified damage to the sterile packaging (blister).Sterility had not been compromised.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Reported event is not related to a combination of products; therefore, a compatibility review is not applicable.The reported event did not occur in an operating room or as part of a medical procedure; medical records are not available for review.The condition of the device when it left zimmer biomet is considered conforming to specification.The root cause of the reported event can be attributed to transit damage and a packaging design issue.The reported event has been confirmed by evaluation of the returned product and provided photos.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Search Alerts/Recalls
|