Additional information provided in d.9., h.3., h.6.And h.11.The product was returned for analysis and broken half optic and haptic was observed.The remaining iol was not returned.The device was received loose in the carton.The lock-out assembly has been removed.Viscoelastic is observed in the device.The plunger has been fully advanced outside the nozzle.No damage observed.Half an optic with haptic was returned outside of the device in a zip lock bag.Solution is dried on the iol.The remaining haptic and optic not returned.The product investigation could not identify the root cause for the reported complaint "doctor notices it was already cracked , haptic broken/torn/missing" as the partial lens was returned outside of the device.Due to this, we are unable to confirm the lens and the plunger position for advancement and determine a root cause.The lens was returned split in two, which is consistent with lens having been explanted.Lens damage may occur: ¿ due to the use of a non-qualified viscoelastic.Lens delivery performance may be negatively affected when using other non-qualified viscoelastic leading to underfill, overfill, misfolding , delivery issues and /or damage.¿ if inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to advance incorrectly and cause delivery issues and /or damage.¿ if the operating room temperature is too high (> 23°c / 73° f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement and cause delivery issues and / or product damage.Any of the above listed causes alone, or in combination, may create the reported event.In addition to this, all iols are 100% cosmetically inspected as per approved manufacturing procedures and the observed lens damage would not meet our current release criteria.Based on these investigation findings, we are unable to verify if the iol/device contributed to the event.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
|