Model Number BR1054096
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2025
Type of Reportable Event Malfunction
Event or Problem Description
A ventilator was returned to a third-party service center for routine preventative maintenance.There was no allegation of device malfunction.The device was not reported to be in patient use.During the evaluation of the device at the third-party service center, an error code related to a permanent failure of the internal battery fault was observed.The technician recommended replacing the device's motor blower assembly to address the issue.No further investigation is required at this time.Additionally, the device was visually inspected and found no evidence of foam degradation.During the evaluation, no foam particles were observed.The device has been remediated; the software has been upgraded and passed all final testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 100
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key21864839
Report Number2518422-2025-104444
Device Sequence Number7898647
Product Code CBK
UDI-Device Identifier00606959022775
UDI-Public(01)00606959022775
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K083526
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberBR1054096
Device Catalogue NumberBR1054096
Device Lot Number180209
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/24/2025
Initial Report FDA Received Date04/18/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/09/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-