| Model Number |
602 |
| Medical Device Problem Codes |
Off-Label Use (1494); Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
Visual Impairment (2138)
|
| Date of Event |
04/03/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was stated by the customer that they are using the yamane technique to implant the lenses.This technique induces a larger amount of force while trying to position the haptics in the scleral tunnel.Our lenses are intended to be implanted in the capsular bag.Thus, the lens was being used off-label.The device history records were reviewed and there were no non-conformities or deviations noted during the manufacturing of the lens that would have caused or contributed to the nature of this complaint.Additionally, our lenses are 100% inspected before they leave our manufacturing site.Therefore, we are confident that the lens was processed per standard operating procedures and inspections and have met all criteria for release.
|
| |
|
Event or Problem Description
|
|
It was reported that the haptic came off.The lens was implanted and then explanted after the haptic broke.There was patient contact, but no harm reported.The procedure was completed with a backup lens.The surgeon used the yamane technique to implant into sclera.
|
| |
|