|
The device, along with the user-provided vest and hose, was returned and evaluated.Functional testing, including external surface temperature measurement, was conducted with the user-supplied components.The evaluation confirmed that the maximum external surface temperature remained within established safety thresholds as defined by iec 60601-1 medical electrical equipment part 1: general requirements for basic safety and essential performance.The reported overheating condition could not be duplicated during testing.
|