| Model Number |
0684-00-0296-02U |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
03/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated fields: b4; d9; g3; g6; h2; h3; h6 type of investigation, investigation findings, investigation conclusion, component code; h11.Corrected fields: d1; d4 model number, lot number, catalog number, udi number no decon or visual inspection performed since product was not returned and could not be evaluated.Therefore, we were unable to confirm or determine a root cause for the reported failure.The failure mode is addressed in the risk file and is operating within its risk profile.The ifu addresses the reported failure.There were no ncmrs identified which could cause or contribute to the reported failure.The investigation does not indicate that the device was inadvertently released as non-conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend (increase in number of complaints over past three (3) months).Based on the rational provided above, no escalation to the capa process is required.
|
| |
|
Event or Problem Description
|
|
It was reported that during intra aortic balloon (iab) therapy the balloon membrane was loosened and insertion was failed.
|
| |
|
Additional Manufacturer Narrative
|
|
Due to character restrictions in block e1 full event site name is (b)(6).Full event site address (b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id (b)(4).
|
| |
|
Search Alerts/Recalls
|