|
It was reported during inspection by the customer that cardiosave intra-aortic balloon pump (iabp) had an issue with electrical test failure, error 53 that diagnose x1 transducer has a value below that stated in the manual, approximately 50.Patient involvement is unknown.
|
|
Updated fields - b4, d9, e2, e3, g3, g6, h2, h3, h6 (type of investigation, investigation findings, investigation conclusions), h11.Corrected field: b5.Added additional medical device problem code.A getinge field service engineer (fse) was dispatched to evaluate the unit.When inspecting the equipment, it was possible to diagnose that the x1 transducer has a value below that stated in the manual, approximately 50.According to the service manual, the value should be close to 750.
|