• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM Back to Search Results
Model Number CAR-170-C
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hypersensitivity/Allergic reaction (1907); Local Reaction (2035); Urticaria (2278)
Date of Event 04/04/2025
Type of Reportable Event Serious Injury
Event or Problem Description
A report was received on (b)(6) 2025 from the nurse of a 56-year-old male patient with a medical history including end stage renal disease, who stated the patient was itchy approximately one hour into a hemodialysis treatment on (b)(6) 2025.Additional information was received on (b)(6) 2025 from the nurse who stated the patient also experienced hives and was administered 50mg intravenous diphenhydramine (benadryl).Per the nurse, the patient continues to treat with nxstage.
 
Additional Manufacturer Narrative
A review of the device history record (dhr) was conducted which confirmed that the device met all quality criteria and manufacturing specifications prior to release.The event is consistent with a hypersensitivity type reaction.Hypersensitivity or allergic adverse reactions are known risks of hemodialysis.The nxstage user guide and instructions for use include allergic reaction as a potential risk associated with dialysis therapy and also include warnings to monitor for potential allergic reactions.Biocompatibility of the device has been established.The patient experienced an event consistent with a hypersensitivity reaction with the same device model number on one additional date.The event on 31 mar 2025 is documented in mfr report # 3003464075-2025-00066.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NXSTAGE SYSTEM ONE
Common Device Name
HIGH PERMEABILITY HEMODIALYSIS SYSTEM
Manufacturer (Section D)
NXSTAGE MEDICAL, INC.
350 merrimack street
lawrence MA 01843
Manufacturer (Section G)
MEDIMEXICO S. DE R. L. DE C. V
av. valle imperial no. 10523
parque industrial valle sur
tijuana 22180
MX   22180
Manufacturer Contact
paula rogalski
nxstage medical, inc.
350 merrimack street
lawrence, MA 01843
9784505276
MDR Report Key21872242
Report Number3003464075-2025-00068
Device Sequence Number18072316
Product Code KDI
UDI-Device IdentifierM535CAR170C0
UDI-Public+M535CAR170C0/$$0826501772303
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K140526
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 04/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/01/2026
Device Model NumberCAR-170-C
Device Catalogue NumberCAR-170-C
Device Lot Number50177230
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/04/2025
Initial Report FDA Received Date04/21/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/25/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
Patient SexMale
-
-