EMBECTA PENEL LIMITED - DUN LAOGHAIRE, IRELAND PEN NDL 31GA 5MM 100BX 1200 USA; Needle, hypodermic, single lumen Back to Search Results
Model Number 320119
Medical Device Problem Codes Misconnection (1399); Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This medwatch submission is both an initial and supplemental filing.Investigation of the results can be seen below: investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed, and no non-conformances were raised in association with this type of event for this lot.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time.
 
Event or Problem Description
Consumer reported that the needles will not attach and the needles are difficult to remove after injection.Consumer does not re-use.Lot #: 4107078 catalog #: 320119 date of event: unknown samples: discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEN NDL 31GA 5MM 100BX 1200 USA
Common Device Name
Needle, hypodermic, single lumen
Manufacturer (Section D)
EMBECTA PENEL LIMITED - DUN LAOGHAIRE, IRELAND
EI 
Manufacturer Contact
avital merl
300 kimball dr.
parsippany, NJ 07054
MDR Report Key21872724
Report Number3023359743-2025-00300
Device Sequence Number7719098
Product Code FMI
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K213478
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial,Followup
Report Date (Section B) 09/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number320119
Device Lot Number4107078
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/26/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/21/2025
Supplement Report FDA Received Date09/08/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
Patient RaceWhite
-
-