STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS FEM COMPONENT CEMENTED; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
|
Back to Search Results |
|
| Catalog Number |
5515-F-401 |
| Medical Device Problem Code |
Malposition of Device (2616)
|
| Health Effect - Clinical Code |
Ambulation or Postural Difficulties (2544)
|
| Date of Event |
03/25/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding malposition involving a triathlon femoral component was reported.The event was not confirmed method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusion: the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
As reported: "rev ltka- for loose tibia and mal alignment." a ps knee was converted to a ts knee with stems and augments.Rep confirmed that no further information will be released by the hospital or surgeon.
|
| |
|
Search Alerts/Recalls
|
|
|