| Model Number |
DTMA1D1 |
| Medical Device Problem Codes |
Pacing Problem (1439); Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
Ventricular Fibrillation (2130)
|
| Date of Event |
04/03/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the cardiac resynchronization therapy defibrillator (crt-d) may have delivered anti-tachycardia pacing (atp) for supra ventricular tachycardia (svt) which advanced to ventricular fibrillation (vf).It was also noted that the right atrial (ra) lead exhibited undersensing potentially triggering the device to falsely classify atrial tachycardia/atrial fibrillation (at/af) episodes as terminated.The crt-d and ra lead remain in use.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: 429688 lead implanted: (b)(6) 2013.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|