| Model Number |
SET-0013-01 |
| Medical Device Problem Codes |
Break (1069); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/14/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The following has been reported: customer reported: running nalaxone drip into primary tubing.End cracked.Patient harm/adverse reaction: no.Stopped active infusion: yes.Resulted in over/under infusion: no.Drug used: nalaxone.Steps taken: got new tubing.A preliminary review identified the following issue: administration set breaks (any part).An active infusion was stopped.Reporting as a conservative measure.No adverse effects were reported.Additional information is needed to complete the investigation.
|
| |
|
Event or Problem Description
|
|
One sample of ivenix administration set was returned for evaluation.No defects were observed during visual inspection.The gold foil corresponded to the specific code set-0013-25.A cracked rotating male luer lock connector was observed at the movil section of the mentioned component.The customer complaint is confirmed.The potential root cause could be related to a lack of compatibility of the luer lock with alcohol or another assembled component.The potential root cause could also be related to improper use by the customer where they exerted too much force on the assembly or applied too much alcohol to the connections causing breakage.An internal investigation was opened to determine the exact root cause.The current process controls detection include 1) post sterilization sampling final inspection, 2) the first piece inspection will be performed to the first final assembled kit manufactured per shift.This kit will be submitted to visual inspection as per the classification of defects included in this specification and as per the applicable drawing, 3) sampling visual inspection is performed for rotating luer lock at incoming inspection area.
|
| |
|
Search Alerts/Recalls
|