According to the complainant the device will not be returned for investigation.We are unable to determine if any product condition could have contributed to the skin irritation.No lot release records were reviewed, as the product lot number was not provided.The omnipod is iso10993 compliant, undergoing cytotoxicity, sensitization, genotoxicity, hemolysis, irritation or intracutaneously reactivity, systemic toxicity, and implantation effects testing.The omnipod is sterilized with 100% ethylene oxide gas with a sterility assurance level of 10-6 per iso11135 and eo residual levels in compliance with iso10993.Each lot is confirmed to meet requirements for non-pyrogen city per iso10993 and sterility per iso11135 prior to release.*please note, that section d is capturing the device identifiers as reported by the complainant.This may not align to the device configuration reported in h11, as this data is pulled from our cloud based on the reported date of event.Cloud - locked down/smartphone lockdown cloud - omnipod 5 software app version 3.1.1 cloud - smartphone operating system n5004l-am-q-mv01602-06-01.06 cloud - smartphone hardware n5004l cloud - cgm sensor type g7.
|