FRESENIUS KABI USA, LLC IVENIX, INC.; LVP PRIMARY ADMINISTRATION SET, SINGLE-INLET, LOW-SORBING Back to Search Results
Model Number SET-0021-1
Medical Device Problem Codes Leak/Splash (1354); No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/02/2025
Type of Reportable Event Malfunction
Event or Problem Description
The following has been reported: nurse reported: cassette leaking from the bottom, gold squares are also not aligned.Patient had dexmedetomidine infusing.Rn noticed moderate size puddle on the floor.Source of leak was noted to be coming from iv tubing cassette.Pump did not alarm for any malfunction.New medication and tubing hung.Old tubing given to cns.Result: needed to get different tubing.- process occurred: before treatment.- injury/adverse reaction: no.- stop active infusion: no.- resulted in under/over-infusion: no.- drug used for infusion: not sure.A preliminary review of the logs identified the following issue: administration set leak (external part).An active infusion was not stopped.Reporting as a conservative measure.No adverse effects were reported.Additional information is needed to complete the investigation.
 
Event or Problem Description
One sample of an ivenix administration set was returned for evaluation.No defects were observed during visual inspection per the applicable drawing.The gold foil were as corresponded to the specific code (b)(6).Primary line infusion passed successfully and was tested with a set rate of 0.5ml/hr and set volume of 0.1ml and primary line was connected to a saline solution bag.An underwater leak test was performed and a leak was detected; bubbles were observed coming from the flow dial.During the microscopic examination a bad welding was revealed between the retainer and seal.The customer complaint is confirmed.The reported leak could be related to poor welding of the cassette.An internal investigation was opened to mitigate the reported issue.The current process controls detection include 1) 100% leak and occlusion test is performed through the leak and occlusion test machine in order to capture units with any blockage or leak failure mode, 2) quality sampling for final physical testing, for performing a flow and underwater leak tests, 3) 100% visual inspection during the manufacturing process and final physical inspections.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IVENIX, INC.
Common Device Name
LVP PRIMARY ADMINISTRATION SET, SINGLE-INLET, LOW-SORBING
Manufacturer (Section D)
FRESENIUS KABI USA, LLC
50 high street
suite 50
north andover MA 01845
Manufacturer (Section G)
FRESENIUS KABI LLC
50 high street
suite 50
north andover MA 01845
Manufacturer Contact
rebecca mccandless
3 corporate drive
lake zurich, IL 60047-8930
MDR Report Key21876709
Report Number3014732157-2025-00352
Device Sequence Number7766095
Product Code FPA
UDI-Device Identifier00811505030108
UDI-Public00811505030108
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K183311
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberSET-0021-1
Device Catalogue NumberSET-0021-1
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/03/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/21/2025
Supplement Report FDA Received Date05/21/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-