| Catalog Number |
212522 |
| Medical Device Problem Code |
Device Markings/Labelling Problem (2911)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported prior to using bd bbl¿ tb stain kit k that one bottle of tb carbolfushin had an expiration date prior to the kit expiration.There was no health impact or consequence reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported prior to using bd bbl¿ tb stain kit k that one bottle of tb carbolfushin had an expiration date prior to the kit expiration.There was no health impact or consequence reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary:
this memo is to summarize findings on your complaint related to kit tb stain kit k, catalog number 212522, batch number 4191205, complaint # (b)(4) for unsatisfactory packaging.Components are mixed and dispensed into the appropriate containers.After qc testing product is released and transported to the distribution center.The batch history record review indicated no discrepancies.All release testing was satisfactory.Complaint trends were reviewed for a period covering 12 months.During that time there has been a trend for this defect identified.The customer reported the kit expiration date is 30-apr2025 but a component within the kit expires 30-sep-2024.This complaint is confirmed based on customer photos and retained kits.A corrective and preventative action (capa) was initiated for the defect in question.Bd will continue to trend on labeling complaints and assess product intended uses against regulatory requirements.If you have further questions, please contact bd technical services.
|
| |
|
Search Alerts/Recalls
|