| Model Number |
173052 |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/31/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10: concomitant product: 173052, 173052 endo uni 12 with (lot#: p4k1067).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a laparoscopic roux-en-y gastric bypass procedure, the device was able to fire 2¿3 staples at the beginning.However, afterward, the handle became ¿blocked¿ unable to open or close and therefore could not fire any more staples.The same issue occurred with another stapler.The issue was resolved by using a new endo universal stapler, which functioned correctly.No patient complications have been reported as a result of this event.
|
| |
|
Event or Problem Description
|
|
It was reported that during a laparoscopic roux-en-y gastric bypass procedure, the device was able to fire 2¿3 staples at the beginning.However, afterward, the handle became ¿blocked¿ unable to open or close and therefore, could not fire any more staples.The same issue occurred with another stapler.The issue was resolved by using a new endo universal stapler, which functioned correctly.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the product noted the instrument was without a sulu.The handle was damaged and broken off.Shearing observed around the handle lugs.The rotational knob was intact and functioned within specification.The shaft collar was intact and functioned within specification.It was reported that the instrument did not fire.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur if the instrument handle is not completely cycled affecting the timing of the instrument creating a jammed condition, and then cycling the handle a second time with force causing the handle to disengage.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: to fire the stapler, simultaneously press the instrument nose against the tissue, mesh or patch and squeeze the handle as far as it will go, then release.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|