The reported adverse event is associated with a returned device; however, the provided clinical information was reviewed by the manufacturer and no specific device analysis is deemed necessary at this time.Previous product examinations have not showed any relationship between a product geometrical deviation and the reported clinical complication.Additionally, there are no indications that a product failure has contributed to the reported issue.Adverse skin reactions around the baha abutment, ranging from slight redness to infected tissue, are common complications with baha implants, and can occur at any time following implantation.The report frequency for these complications is being monitored under cbas complaint and medical device reporting data monitoring plan and the status is updated on a monthly basis.This event is added to this monitoring.
|
|
Per the clinic, the patient experienced infection at the implant site.Subsequently, the patient was treated with multiple courses of oral antibiotics (date and duration not reported).The patient also underwent a revision surgery under general anesthesia on (b)(6) 2025 in order to perform skin flap advancement and to irrigate the infected area with antibiotic solution.The device was explanted during the same surgery.
|