| Catalog Number |
130742206 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Wound Dehiscence (1154); Unspecified Infection (1930)
|
| Date of Event |
03/26/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient underwent revision surgery for the left shoulder.Initially the patient underwent reverse shoulder replacement due to a fractured left humerus.A couple of weeks later the patient's joint had dislocated.The glenosphere changed at that time to lateralized +0 mm 42mm standard glenosphere; and pe cup 42+6mm.Following this second operation, the surgical wound dehisced and looks infected.The patient went back to theatre and had specimens taken and the joint washed out.The glenosphere and cup were removed.New glenosphere 42mm and standard 42 +6mm pe cup implanted.No implants to return as retained by hospital.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy orthopaedics is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy orthopaedics has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy orthopaedics or its employees that the report constitutes an admission that the device, depuy orthopaedics, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: d4: udi: as the reported lot number for the device involved in the event is not valid, the full udi is currently not available.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A manufacturing record evaluation (nc search) was performed for the finished device product code: 130742206, lot: 5698730 and no non-conformances / manufacturing irregularities were identified.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: a manufacturing record evaluation (nc search) was performed for the finished device product code: 130742206, lot: 5698730 and no non-conformances / manufacturing irregularities were identified.H11 additional narrative: added: d10.Corrected: d4 (primary udi number).
|
| |
|
Search Alerts/Recalls
|