| Model Number |
CNW0T3 |
| Medical Device Problem Codes |
Material Opacification (1426); Scratched Material (3020)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
03/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A non-healthcare professional reported during an intraocular lens (iol) implant procedure, immediately after the lens implant, the optic surface was cloudy, and scratches appeared, so it was promptly explanted.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided in h.3., h.6.And h.11.The product was not returned for analysis.Only photo was returned.The reported complaint is confirmed from the returned photo.Provided photo shows an implanted iol.There appears to be cloudiness/white marks and light reflections over the iol optic.The exact location of the marks ( anterior or posterior surface) cannot be confirmed from the returned photo.The reported cloudiness may potentially be related to polychromatic sheen.The root cause is deemed to be manufacturing related.The cause of ¿sheen¿ or ¿film-like¿ appearance is a result of a change in the positioning of the iol during the manufacturing process and is not associated with any foreign material or contamination.The observation has been reported clinically and confirmed in laboratory settings to rapidly disappear following implantation and should no longer be visible, or very little remaining, by the one-day postoperative visit.The appearance is a result of a temporary change in the surface of the iol that occurs during the delivery process which is only visible under certain orientations and illumination settings.This quickly resolves once the iol is delivered and exposed to intraocular temperature over time.There are no adverse impacts or lasting effects on the iol or the patient, and no adverse events or clinical impact on vision have been reported associated with this finding.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|