Catalog Number 151810035
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Fibrosis (3167); Joint Laxity (4526); Swelling/ Edema (4577)
Date of Event 01/29/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy orthopaedics is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy orthopaedics has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy orthopaedics or its employees that the report constitutes an admission that the device, depuy orthopaedics, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: h3, h6: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.No device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
On (b)(6) 2023 patient had underwent right knee with velys.On (b)(6) 2025, patient underwent a right tka revision due to pain, swelling, and instability.Patient was found to have aseptic femoral and tibial loosening at the cement to implant interface.There was underlying fibrous tissue and the cement was removed after the tibial component and femoral components were removed.The patella was noted to have been removed as well, but there was no mention implant loosening related to the patella.Competitor implants were placed at revision.Doi: (b)(6) 2023.Dor: (b)(6) 2025.Affected side: right knee.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE MEDIAL ANAT PAT 35MM
Common Device Name
ATTUNE IMPLANT : KNEE PATELLA
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key21882059
Report Number1818910-2025-06120
Device Sequence Number15611205
Product Code JWH
UDI-Device Identifier10603295056638
UDI-Public(01)10603295056638(17)271130(10)4002367
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number151810035
Device Lot Number4002367
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/26/2025
Initial Report FDA Received Date04/22/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/11/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age59 YR
Patient SexMale
-
-