Catalog Number 0502537030
Medical Device Problem Codes Fogging (1253); Optical Distortion (3000)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/25/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was foggy image.
 
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that there was foggy image.
 
Additional Manufacturer Narrative
The product was returned for investigation and the reported failure mode was confirmed.Alleged failure: scope is blurry and foggy confirmed failure: outer tube damaged (bent, dented) damaged/missing eyepiece out of focus probable root cause: ¿ laser welding seal failure ¿ distal/proximal window solder failure ¿ damage to optical train ¿ damage to needle ¿ damage to distal or proximal windows ¿ moisture intrusion ¿ end of life wear-out ¿ environmental disturbance: endoscope colder than dew point ¿ environmental disturbance: fluid entrapment between the coupler and eyepiece junction (eyepiece only) ¿ use error the failure mode will be monitored for future reoccurrence.E1: initial reporter phone number and initial reporter postal code added.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIM HD LAPAROSCOPE AUTOCLAVABLE 5.4MM X 30DEG 30CM
Common Device Name
CONFOCAL OPTICAL IMAGING
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
chester rebugio
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key21882336
Report Number0002936485-2025-00362
Device Sequence Number15613571
Product Code OWN
UDI-Device Identifier07613327059373
UDI-Public07613327059373
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K142310
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0502537030
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/25/2025
Supplement Date Received by Manufacturer06/11/2025
Initial Report FDA Received Date04/22/2025
Supplement Report FDA Received Date07/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-