| Model Number |
0684-00-0479-01 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Due to character restrictions in block e1 full event site address is (b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that intra aortic balloon did not inflate after insertion.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated fields: b4;d9; e2; e3; g3; g6; h2; h3; h6 type of investigation, investigation findings, investigation conclusion, component code; h11.Corrected fields: d1; d4 model number, catalog number, udi number, expiration date, h4.No decon or visual inspection performed since product was not returned and could not be evaluated.Therefore, we were unable to confirm or determine a root cause for the reported failure.The failure mode is addressed in the risk file and is operating within its risk profile.The ifu addresses the reported failure.There were no ncmrs identified which could cause or contribute to the reported failure.The investigation does not indicate that the device was inadvertently released as non-conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend (increase in number of complaints over past three (3) months).Based on the rational provided above, no escalation to the capa process is required.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected fields: h6 health effect ¿ impact codes, health effect ¿ clinical code.Updated fields: b4; g3; g6; h2; h11.
|
| |
|
Additional Manufacturer Narrative
|
|
Correceted field: b5.Updated field: b4; g3; g6; h2; h11.
|
| |
|
Event or Problem Description
|
|
It was reported that intra aortic balloon did not inflate after insertion.There was no injury or harm reported.
|
| |
|
Search Alerts/Recalls
|