| Model Number |
BF-P190 |
| Medical Device Problem Code |
Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/26/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the broncho videoscope displayed a b30 error code (scope communication error).The issue occurred during inspection before use.There were no reports of patient involvement.
|
| |
|
Event or Problem Description
|
|
No additional information received from customer.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d9, d10, h2, h3, h6 (type of investigation, investigation findings, investigation conclusions), h11.The device was returned to olympus for inspection and the reported failure was confirmed.A definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the error code malfunction: cause traced to electrical/electronic component problem; expected or random component failure without any design or manufacturing issue.Olympus will continue to monitor field performance for this device.
|
| |
|