ANGIODYNAMICS, INC SMART PORT CT; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Model Number H787CT96STSD0
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code Thrombosis/Thrombus (4440)
Date of Event 03/01/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is not available to be returned to the manufacturer for evaluation.An investigation into the root cause for the event is currently in progress.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Event or Problem Description
On or about (b)(6) 2015, plaintiff underwent placement of the angiodynamics smartport power injectable port, lot number 4777969.The device was implanted by (b)(6), md , at (b)(6) , for the purpose of ongoing vein access.On or about (b)(6) 2018, plaintiff presented himself to (b)(6) where his port was removed due to causing blood clots and an occlusion.
 
Additional Manufacturer Narrative
The customer's reported complaint description of patient experienced thrombosis and catheter occluded cannot be confirmed given the patient centric nature of this serious adverse event (sae).No port product was returned to angiodynamics for evaluation since there was no reported port device malfunction, just patient sae of thrombosis.Thrombosis/clot formation/occlusion is cautioned in the device dfu as potential complication for an implanted port system.A device history record (dhr) review of the indicated packaging lot revealed no quality related issues or manufacturing deficiencies at the time of manufacture.Labeling review: the directions for use (dfu) that is supplied with this port device contains the following statements: warnings: - absence of a blood return or a poor blood return can be a sign of a potential complication such as occlusion, kinking, breakage, pinch-off syndrome, fibrin formation, thrombosis or malposition.This should be evaluated prior to device usage.A blood return should be present prior to usage of device for any therapy or testing.- if the patient complains of pain, or if there is swelling when the device is flushed or when medication or contrast media is administered, evaluate the device for infiltration, proper needle placement, and potential complications such as occlusion, kinking, breakage, pinch-off syndrome, thrombosis or malposition.Failure to assess these complaints or observations can lead to device failure.Precautions: · for peripheral placement, irritation to the vein, resulting in postoperative thrombophlebitis, has been associated with guidewire and introducer insertion.· when using percutaneous introducers: to avoid blood vessel damage, do not allow the percutaneous introducer sheath to remain indwelling in the blood vessel without the internal support of a catheter or dilator.· if more than one drug is to be administered, between the individual drug applications, flush the system with 5 to 10 ml normal saline for injection to prevent drug interactions.· after any infusion, injection or bolus application, the system should be flushed with normal saline for injection or locked with a heparin solution per institutional protocol to prevent thrombotic occlusion of the catheter.Potential complications: use of an angiodynamics port system involves potential risks normally associated with the insertion or use of any implanted device or indwelling catheter including but not limited to: these complications are well documented in literature and should be considered when a venous access device is utilized.Air embolism.Clot formation.Fibrin sheath.Infection.Inflammation.Occlusion.Thromboembolism.Thrombophlebitis.Thrombosis.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint continues to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMART PORT CT
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
ANGIODYNAMICS, INC
10 glens falls technical park
glens falls NY 12801
Manufacturer (Section G)
ANGIODYNAMICS, INC
10 glens falls technical park
glens falls NY 12801
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587990
MDR Report Key21883517
Report Number1317056-2025-00106
Device Sequence Number20808275
Product Code LJT
UDI-Device Identifier15051684018067
UDI-Public15051684018067
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K062414
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2018
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2017
Device Model NumberH787CT96STSD0
Device Catalogue NumberCT96STSD
Device Lot Number4777969
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/02/2025
Supplement Date Received by Manufacturer05/21/2025
Initial Report FDA Received Date04/22/2025
Supplement Report FDA Received Date05/22/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/25/2014
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other; Hospitalization;
Patient SexMale
-
-