Model Number CD3365-40Q
Medical Device Problem Code Interrogation Problem (4017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/02/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a lack of radio frequency (rf) telemetry communication was observed between the implantable cardioverter defibrillator and programmer.A cybersecurity firmware update was performed, and the issue was resolved.There were no reported patient symptoms.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA CRT-D QUAD RF HV
Common Device Name
No Match
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key21883836
Report Number2017865-2025-50794
Device Sequence Number3442199
Product Code NIK
UDI-Device Identifier05414734508292
UDI-Public(01)05414734508292(10)A000017617(17)180131
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P030054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 04/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2018
Device Model NumberCD3365-40Q
Device Catalogue NumberCD3365-40Q
Device Lot NumberA000017617
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/02/2025
Initial Report FDA Received Date04/22/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/04/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MED 6935M-62; QUARTET; TENDRIL STS
Patient SexMale
-
-