Customer reported patient presents with necrotic bone, purulence, redness and swelling, tooth 19.I&d (incision and drainage) external chon drain placed and removal of implant site #19 and debridement.Symptoms as a result of the event: pain, inflammation.Additional appointment required: yes, possible replacement.
|
|
Zimvie complaint number (b)(4).The following fields have been updated: b4: date of this report.B5: describe event or problem.G3: date received by manufacturer.G6: type of report.H1: type of reportable event.H2: follow up type.H3: device evaluated by manufacturer.H6: adverse event problem.H10: additional narrative.Zimvie received one (1) xiitp4313, (osseotite tapered certain prevail implant 4/3 x 13mm) x 10mm (l) for evaluation.Visual evaluation was performed, the implant had signs of use with debris on its external threads.The implant had some removal damage near the platform area.However, there was no damage identified or signs of malfunction that would have contributed to the event.Measurements match drawing.Dhr review was completed for the subject lot number 2120016476.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2120016476 for similar events and no other complaint was identified.Review completed utilizing keywords: dental : medical : infection and dental : medical : other.Based on the investigation and risk management file review, the most likely root cause determined from the investigation was patient factors.Therefore, based on the available information, a device malfunction did not occur.No damage to the implant was identified.The reported event was non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.Similar complaints for implant infection have been previously investigated.Visual and dimensional evaluations of the previously returned product have not identified or suggested manufacturing non-conformances.While non-conformances were identified for some lots during manufacturing record reviews, the documented disposition actions for each did not suggest the likely release of non-conforming product.Zimvie quality management system (qms) has controls in place to ensure the distribution of conforming product.Additionally, all device history record reviews verified that each implant was sterilized per procedure for every device.All complaint data used for the summary investigation was found to be conforming and did not meet capa/hhe/d/ or any further escalations.Therefore, there were no complaints which confirmed a manufacturing or design related issue that did or could cause or contribute to the reported event.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
|