• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I OSSEOTITE® TAPERED CERTAIN® PREVAIL® IMPLANT 4/3 X 13MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I OSSEOTITE® TAPERED CERTAIN® PREVAIL® IMPLANT 4/3 X 13MM; DENTAL IMPLANT Back to Search Results
Catalog Number XIITP4313
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Necrosis (1971)
Date of Event 03/06/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A4: weight unknown / not provided.
 
Event or Problem Description
Customer reported patient presents with necrotic bone, purulence, redness and swelling, tooth 19.I&d (incision and drainage) external chon drain placed and removal of implant site #19 and debridement.Symptoms as a result of the event: pain, inflammation.Additional appointment required: yes, possible replacement.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).The following fields have been updated: b4: date of this report.B5: describe event or problem.G3: date received by manufacturer.G6: type of report.H1: type of reportable event.H2: follow up type.H3: device evaluated by manufacturer.H6: adverse event problem.H10: additional narrative.Zimvie received one (1) xiitp4313, (osseotite tapered certain prevail implant 4/3 x 13mm) x 10mm (l) for evaluation.Visual evaluation was performed, the implant had signs of use with debris on its external threads.The implant had some removal damage near the platform area.However, there was no damage identified or signs of malfunction that would have contributed to the event.Measurements match drawing.Dhr review was completed for the subject lot number 2120016476.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2120016476 for similar events and no other complaint was identified.Review completed utilizing keywords: dental : medical : infection and dental : medical : other.Based on the investigation and risk management file review, the most likely root cause determined from the investigation was patient factors.Therefore, based on the available information, a device malfunction did not occur.No damage to the implant was identified.The reported event was non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.Similar complaints for implant infection have been previously investigated.Visual and dimensional evaluations of the previously returned product have not identified or suggested manufacturing non-conformances.While non-conformances were identified for some lots during manufacturing record reviews, the documented disposition actions for each did not suggest the likely release of non-conforming product.Zimvie quality management system (qms) has controls in place to ensure the distribution of conforming product.Additionally, all device history record reviews verified that each implant was sterilized per procedure for every device.All complaint data used for the summary investigation was found to be conforming and did not meet capa/hhe/d/ or any further escalations.Therefore, there were no complaints which confirmed a manufacturing or design related issue that did or could cause or contribute to the reported event.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE® TAPERED CERTAIN® PREVAIL® IMPLANT 4/3 X 13MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key21884601
Report Number0001038806-2025-00850
Device Sequence Number3442225
Product Code DZE
UDI-Device Identifier00844868022138
UDI-Public(01)00844868022138(17)290704(10)2120016476
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K130949
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 07/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue NumberXIITP4313
Device Lot Number2120016476
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/22/2025
Supplement Report FDA Received Date07/22/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/04/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age84 YR
Patient SexFemale
Patient EthnicityNon Hispanic
-
-