|
This follow-up report is being submitted to relay additional and corrected information.The following sections were updated: b4, d9, g3, g6, h2, h3, h6, h11.The following sections were corrected: d4, h4.Visual evaluation of the returned product identified signs of use and implantation (gouges/surface scuff marks).The locking bar clasp was also confirmed to be bent away from the body of the implant.Dimensional analysis of the product found that one of the product features was out of print specifications, likely caused from insertion, explanation and damage seen on the clasp.The device history records were reviewed and no discrepancies were identified.A definitive root cause could not be determined with the information provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|