| Catalog Number |
SD800.435 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Unspecified Mental, Emotional or Behavioural Problem (4430)
|
| Date of Event |
01/01/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient underwent removal surgery due to infection.Implant 1: in 2018, after the first intervention, the patient suffered a liver infection that eventually migrated to the brain and therefore had to remove the implant for safety.Implant 2: in 2023, after the second operation, the patient had another infection in the area of the implant which advised the removal of the implant again.Currently, the patient suffers from craniectomized syndrome, that is to say, the brain has no protective bone and is very sensitive to its own atmospheric pressure which prevents it from living a normal life.The patient can only lie down without suffering from dizziness, so it is indicated a new plastia that allows him to return to normal life.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #:(b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11: additional narrative: h3, h4, h6: part number: sd800.435, lot number: 7848p62: manufacturing site: mezzovico.Release to warehouse date: september 12, 2023.A manufacturing record evaluation was performed for the not sterile finished lot number, and no non-conformances were identified.H3, h6: the photo investigation revealed nothing indicative of a device nonconformance.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was not confirmed as the observed condition of the psi sd800.435 peek implant would have not contributed to the reported adverse event.¿based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: corrected: h6 (health effect - clinical code).H3, h6: the product was not returned to j&j medtech orthopaedics, however photos were provided for review.The photo investigation revealed nothing indicative of a device nonconformance.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was not confirmed as the observed condition of the psi sd800.435 peek implant would have not contributed to the reported adverse event.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review: part number: sd800.435.Lot number: 7848p62.Manufacturing site: mezzovico.Release to warehouse date: 12 sep 2023.A manufacturing record evaluation was performed for the not sterile finished lot number, and no non-conformances were identified.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|