|
"a number of risk factors for rupture have been identified; the most common cause is surgical instrument damage.¿ (handel, garcía and winxtrom, 2013.Breast implant rupture: causes, incidence, clinical impact, and management.Plast.Reconstr.Surg.132: 1128.).¿it is imperative that careful planning and exacting surgical technique are used to reduce iatrogenic damage of implants during primary placement and subsequent reoperations.¿ (handel, garcía and winxtrom, 2013.Breast implant rupture: causes, incidence, clinical impact, and management.Plast.Reconstr.Surg.132: 1128).Breast implants are not considered lifetime devices, and implant rupture is a risk associated with breast surgery, which can occur at any time.Additionally, rupture is a common and known reason for complaints, and it is clearly characterized in the product dfu included with the implant, as stated above.Additionally, no cases of rupture because of product failure have been reported to establishment labs ever.The instructions for use in the directions for use were reviewed to determine if there are indications that ensure the prevention and good handling of the product.For this event, it is considered that the information is suitable on the section of surgical precautions as follows: ¿breast implants can potentially remain intact for decades in the body, but all such devices will fail at some point.Breast implants rupture when the shell develops a tear or hole.Rupture can occur at any time after implantation, but it is most likely to occur; the longer the implant is in place.The following may cause implants to rupture: damage by surgical instruments, implant stress, and weakening during implantation, age and design of the implant, submuscular rather than sub-glandular location, the occurrence of post-operatory hematomas or seromas, folding or wrinkling of the implant shell, excessive force to the chest, trauma, compression during mammographic imaging, and severe capsular contracture¿.Also, a complete review of the dhr's was carried out, no deviation was found in the process, so it is established that there is no evidence of a relationship between the event and the manufacturing process.Device involved is in the lab pending for testing.
|
|
Korea - it was reported that the patient visited the clinic due breast pain, the examination revealed a defect in the anterior shell of the left implant indicating rupture.It was discovered during routine check up using the ultrasound.No patient demographics, such as age, weight, or ethnicity, were provided by the reporter, and no adverse outcomes for the patient have been noted.Efforts to gather additional information are ongoing, and the investigation remains in progress.
|
|
General conclusion: device involvement: a causal relationship between the reported event and the device has been established.Event characterization: the event was classified as a device rupture associated with the use of the insertion sleeve.Technical investigation summary: laboratory testing: laboratory evaluation was conducted in accordance with wi-001048 "pms laboratory testing units." key findings include: visual inspection: during the visual inspection of the unit, a device rupture was identified.The rupture displayed a circular pattern, similar to those observed in the lt-001474 laboratory test for motiva® insertion sleeve functional test.Microscopic analysis: microscopic examination revealed marks on the shell, these marks differ from those typically caused by spontaneous tears in the implant shell or cutting marks (surgical damage), as outlined in the sid-001085 "visual aid for cutting marks inspection." shell compliance testing: mechanical and material integrity testing confirmed that the implant shell met all applicable international specification standards.Test result analysis: the performance results obtained from the device involved in the complaint correlate with the outcomes observed during the lt-001474 motiva® insertion sleeve functional test, identified as part of the most assignable cause investigation.The lt demonstrated proper device performance under condition a (dfu-compliant use), with all samples completing multiple insertion cycles without damage.Conversely, conditions b and c (non-dfu-compliant use) conditions led to consistent implant damage and insertion failures.These results correlate with the device behavior observed in the complaint, supporting that the event was caused by deviation from the instructions for use rather than a device defect or design issue.Device history record (dhr) review: a comprehensive review of the device history record (dhr) for the affected lot was completed.No deviations, non-conformances, or anomalies were identified during manufacturing.All raw materials and process parameters conformed to specified quality requirements.Trend and signal monitoring: the event has been evaluated as part of ongoing trend analysis activities within the pms program.Per the current complaint data report: no adverse or unexpected trends related to the alleged event have been identified.No further corrective or preventive actions are required at this time.Establishment labs will continue to monitor for similar events as part of routine pms surveillance.
|