• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. PDSII CLR 18IN 4-0 S/A PC-3 PRM MP; SUTURE, SURGICAL, ABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. PDSII CLR 18IN 4-0 S/A PC-3 PRM MP; SUTURE, SURGICAL, ABSORBABLE Back to Search Results
Catalog Number Z845G
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an unknown surgery on an unknown date and suture was used.During the procedure, it was reported by the distributor per the account that product is breaking/snapping.No patient harm was reported.Product is available for return.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint#: (b)(4).H6: component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformance's were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The following information was requested, but unavailable: the only information i have: when was the suture broke/snapped (in the package, during removal from package, during handling prior to use on patient or during use on the patient)? please specify.How many devices demonstrated the alleged deficiency? did the event occur during one or multiple patient procedures? what is the total number of procedures? have any of these events been previously reported to ethicon? if so, could you please provide the corresponding reference number(s)? could you kindly perform the follow-up attempt for the product return? please ensure that the shipment tracking number is provided.Please provide the source or name of person providing answers to follow-up questions.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 5/20/2025 this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H3 investigational summary: the product was returned to ethicon for evaluation.One open box with nine representative unopened samples was received for analysis.Upon initial inspection of the samples, no external damages were observed on the external packets.In order to evaluate the condition of the returned samples, the packets were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related to the customer complaint and no defects were observed during evaluation.The functional test was performed using instron equipment and the tensile force was above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PDSII CLR 18IN 4-0 S/A PC-3 PRM MP
Common Device Name
SUTURE, SURGICAL, ABSORBABLE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key21885715
Report Number2210968-2025-04404
Device Sequence Number8146770
Product Code NEW
UDI-Device Identifier10705031061866
UDI-Public(01)10705031061866(17)47208(10)100BZ1
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
N18331
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberZ845G
Device Lot Number100BZ1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/21/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/24/2025
Supplement Date Received by Manufacturer04/30/2025
Initial Report FDA Received Date04/22/2025
Supplement Report FDA Received Date05/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/08/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-