|
The most common overall indication for reoperation is capsular contracture.(handel, 2006).Breast prostheses are no different from any foreign material implanted into the human body in the sense of triggering a protective immune reaction from the host.This ¿foreign body response¿ (fbr) is universal and ideally removes or otherwise surrounds the ¿irritant material¿ with fibrous tissue to prevent unwanted immune sequelae.A capsule around a breast implant is, therefore, a necessary mechanism of body defense, but if excessive it can lead to pain and deformity of the breast (steiert, et al., 2013).It has been identified several significant risk factors for capsular contracture, including device features (surface, size), surgical factors (periareolar incision, subglandular placement, antibiotic irrigation), the development of hematoma/seroma, and the use of a surgical bra.(calobrace, 2018).Periprosthetic capsules become pathologically active and undergo a ¿constrictive fibrosis¿ due to retraction of the fibrous tissue, which deforms their contents and impairs the aesthetic outcome, manifested by hardening of variable degree and, in advanced cases, by deformity of the breast.(baker et al., 1990).Per our directions for use, each patient must receive the establishment labs s.A.¿motiva implants®: information for the patient¿.During her surgical consultation, it is the surgeons´ responsibility to ensure that the patient completely understands the information regarding the risks, benefits, and recommendations associated with silicone gel-filled breast implant surgery, as well as the complications typical of any type of surgery.This document is available in motiva® website: ifu.Motiva.Health.The instructions for use in the directions for use were reviewed to determine if there are indications that ensure the prevention and good handling of the product.For this event, it is considered that the information is suitable on the section of surgical precautions as follows: ¿capsular contracture occurs when the capsule tightens and squeezes the implant.This can cause the implant to turn rigid (from slightly firm to quite hard) and the firmest ones can cause varying degrees of discomfort, pain, and palpability.In addition to the firmness, capsular contracture can result in a deformed breast, visible surface wrinkling and/or displacement of the implant.Detection of breast cancer by mammography may also be a more difficult.Capsular contracture may be more common following infection, hematoma, and seroma, and the chance of it happening may increase over time.Capsular contracture is a risk factor for implant rupture, and it is the most common reason for reoperation in augmentation and reconstruction patients.Patients should also be advised that additional surgery might be needed in cases where pain and/or firmness are severe (baker grades iii or iv) and that capsular contracture may happen again after additional surgeries.Correction of capsular contracture may require surgical removal or release of the capsule, or removal and possible replacement of the implant itself." also, a complete review of the dhr was carried out, no deviation was found in the process, so it is established that there is no evidence of a relationship between the event and the manufacturing process.General conclusion: the alleged capsular contracture baker grade iii was confirmed according to the clinical evidence provided.The aetiology of capsular contracture is multifactorial and consisting of patient, type of surgery, prosthetic material used, the implant pocket placement, the incision type and the duration of follow-ups (liu et al., 2015) (headon, kasem & mokbel, 2015).Capsular contracture is a risk associated with breast surgery (handel, 2006) and there is no evidence to suggest a link between this particular implant and/or its manufacturing process and a higher risk of capsular contracture occurrence.The cause of capsular contracture is multifactorial (calobrace, 2018), and we cannot conclude that the reported event was caused by the manufacturing process and/or the product itself.
|