• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PRECISION IDEAL EYE 5.5 30° HD ACL LAP30; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PRECISION IDEAL EYE 5.5 30° HD ACL LAP30; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Catalog Number 0502503030
Medical Device Problem Codes Fogging (1253); Moisture or Humidity Problem (2986)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/25/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that there was foggy image.
 
Event or Problem Description
It was reported that there was foggy image.
 
Additional Manufacturer Narrative
According to what was reported, the device was available for evaluation.This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: foggy image probable root cause: ¿ laser welding seal failure ¿ distal/proximal window solder failure ¿ damage to optical train ¿ damage to needle ¿ damage to distal or proximal windows ¿ moisture intrusion ¿ end of life wear-out ¿ environmental disturbance: endoscope colder than dew point ¿ environmental disturbance: fluid entrapment between the coupler and eyepiece junction (eyepiece only) ¿ use error the reported failure mode will be monitored for future reoccurrence.The manufacture date is not known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION IDEAL EYE 5.5 30° HD ACL LAP30
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
chester rebugio
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key21886356
Report Number0002936485-2025-00369
Device Sequence Number3442283
Product Code GCJ
UDI-Device Identifier07613327059144
UDI-Public07613327059144
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K910132
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0502503030
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/25/2025
Supplement Date Received by Manufacturer06/09/2025
Initial Report FDA Received Date04/22/2025
Supplement Report FDA Received Date07/07/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-