• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS 1-PIECE IOL WITH TECNIS SIMPLICITYTM DELIVERY SYSTEM; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS 1-PIECE IOL WITH TECNIS SIMPLICITYTM DELIVERY SYSTEM; INTRAOCULAR LENS Back to Search Results
Model Number DCB00
Medical Device Problem Codes Break (1069); Device Difficult to Setup or Prepare (1487)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/26/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Section a2, a4 and a5: per regulation eu 2016/679 (general data protection regulation), patient identifiers were not collected or recorded and therefore are not available.Section d6a: implant date: n/a - lens was not implanted.Section d6b: explant date: n/a - lens was not implanted.Section e1: telephone number: (b)(6).Device evaluation: at the time of the investigation, device was not available for evaluation, therefore no testing could be performed.The reported event cannot be confirmed.Manufacturing record review: the manufacturing records for the device were reviewed.The product was manufactured and released according to specification.Conclusion: based on the investigation results there is no indication of a product quality deficiency.Attempts have been made to obtain missing information.However, to date, no response has been received.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the plunger in the preloaded, monofocal, intraocular lens (iol) did not function as expected, resulting in the lens being pierced.Consequently, the device was discarded, and a second device was successfully implanted without any delay to the surgery.There was no patient contact with the device and thus, no patient impact.No further information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1-PIECE IOL WITH TECNIS SIMPLICITYTM DELIVERY SYSTEM
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key21886538
Report Number3012236936-2025-000087
Device Sequence Number8189941
Product Code HQL
UDI-Device Identifier05050474636156
UDI-Public(01)05050474636156(17)271010
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDCB00
Device Catalogue NumberDCB0000235
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/27/2025
Initial Report FDA Received Date04/22/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/09/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-