• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ NATURAL SINGLEPIECE IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ NATURAL SINGLEPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number SN60WF
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/13/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
Batch case received from ha portal, reported during an intraocular lens (iol) implant procedure, haptic was found to be broken when preparing to load into the chamber.The same specification and model of iol was immediately replaced.Additional information was requested.
 
Additional Manufacturer Narrative
Additional information provided in h.3., h.6.And h.11.The product was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.The account indicated the use of an unspecified company cartridge and handpiece with a non-company viscoelastic.The root cause for the reported complaint may be related to a failure to follow the ifu.The viscoelastic indicated is not qualified for use with this lens, with any qualified lens/cartridge combination.The ifu instructs that an alcon qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.The initial report indicated the product was not available to evaluate.Due diligence has been performed in an attempt to obtain further information on this event.File will be reopened if new information is received.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ NATURAL SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key21888566
Report Number1119421-2025-01148
Device Sequence Number17004978
Product Code HQL
UDI-Device Identifier00380655093276
UDI-Public(01)00380655093276(17)290827(11)240829(10)15914335
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSN60WF
Device Catalogue NumberSN60WF.240
Device Lot Number15914335
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/27/2025
Supplement Date Received by Manufacturer05/23/2025
Initial Report FDA Received Date04/23/2025
Supplement Report FDA Received Date06/18/2025
Date Device Manufactured08/29/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH IOL DELIVERY SYSTEM, CARTRIDGE; MONARCH IOL DELIVERY SYSTEM, INJECTOR; UNSPECIFIED NON-ALCON VISCOELASTIC
Patient Age86 YR
Patient SexFemale
-
-