SYNTHES GMBH AO/ASIF QUICK COUPLING FOR DRILL BITS; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
| Catalog Number |
05.001.250 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/08/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the ao/asif quick coupling for drill bits device drill head was sheared off.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed, and it was determined that the quick coupling device had a broken drive shaft, and component damage.It was further determined that the device failed pretest for general condition.Therefore, the reported condition was confirmed.The assignable root cause of this condition was determined to be traced to component failure due to wear.
|
| |
|
Search Alerts/Recalls
|
|
|