• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAILLEFER INSTRUMENTS HOLDING SARL HEDSTROEM M-ACCESS 25MM 008; FILE, PULP CANAL, ENDODONTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAILLEFER INSTRUMENTS HOLDING SARL HEDSTROEM M-ACCESS 25MM 008; FILE, PULP CANAL, ENDODONTIC Back to Search Results
Catalog Number A16MA02500812
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Pain (1994)
Date of Event 04/05/2025
Type of Reportable Event Malfunction
Event or Problem Description
In this event it is reported that hedstroem m-access 25mm 008 used in performing treatment to remove the gutta percha points from the root canal, the patient experienced pain due to improper operation causing lateral penetration of the root canal.The patient was given given oral antibiotics and local disinfection has been done.
 
Additional Manufacturer Narrative
Since this event resulted in medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, it must be presumed that the malfunction would be likely to cause or contribute to a serious injury should it recur.As such, this event meets the definition of a reportable event per 21 cfr part 803.The device was not returned for evaluation.However, the lot number was provided and retained-product testing and/or dhr review are planned.The results will be submitted as they become available.
 
Additional Manufacturer Narrative
Summary: involved product that didn't give satisfaction during use was not returned and cannot be analyzed.Moreover, no unused file is available for evaluation.Nothing unusual to report was found during dhr review (batch #1784412).Root causes are not identified.We will track this kind of event and monitor the trend.Potential root causes may be incorrect technique (hand used file - technique no more verifiable to date), overuse (number of uses not communicated), excessive wear, patient condition and behaviour during treatment or material issue (no analysis of the file possible).In addition, there are further circumstances on the conditions in dentistry (used disinfectants, training/knowledge status), or any other environmental conditions, which are unknown to us and may also have an impact on the reported failure mode.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEDSTROEM M-ACCESS 25MM 008
Common Device Name
FILE, PULP CANAL, ENDODONTIC
Manufacturer (Section D)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ  CH-1338
Manufacturer (Section G)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ   CH-1338
Manufacturer Contact
dan eagar
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key21889861
Report Number8031010-2025-00060
Device Sequence Number3405900
Product Code EKS
UDI-Device IdentifierJ003A16MA025008121
UDI-PublicJ003A16MA025008121
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 06/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Catalogue NumberA16MA02500812
Device Lot Number1784412
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date04/14/2025
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/14/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/23/2025
Supplement Report FDA Received Date06/13/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-