Additional information was provided in sections d.9, h.3, h.6 and h.11.Five opened ophthalmic side port knives were received for the report of blunt knife.Samples were visually inspected and all were found to be nonconforming with bent and/or broken and/or curled tip and damaged cutting edge.Penetration testing could not be performed due to the damage of the samples.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The investigation conducted a non-conformance review of the reported lot number.No deviations were identified and all corresponding production release specifications defined in the device master record were met.The returned samples are visually non-conforming with a bent tip and damaged cutting edge.The damage to the returned samples is consistent with damage that can occur when the blade contacts a hard surface such as the protective blade tray when product is improperly removed or inserted after use, or improper handling, or contact with another instrument during surgery or set-up.The damage seen on the returned complaint samples could have contributed to the customers reported issue of blunt knife.Based on the evaluation of the information and materials received, the investigation was unable to identify the root cause or origin of the reported event.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.All knives are 100% inspected by trained operators using a minimum of 10x magnification during manufacturing.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.The manufacturer internal reference number is: (b)(4).
|