MEDOS INTERNATIONAL SàRL CH CONF INTRO NDLE, DIA, 11G 4"; NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE Back to Search Results
Catalog Number 283903411
Medical Device Problem Codes Break (1069); Entrapment of Device (1212); Material Deformation (2976)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 03/26/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during surgery, the needles bent during insertion.When removing the curved needles, the tips broke off.The tips could not be removed from the patient and remained in the patient.There was a surgical delay of ten minutes related to the event.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: the product was returned to j&j medtech orthopaedics for evaluation.Visual analysis of the returned sample revealed that the conf intro ndle, dia, 11g 4" was broken at the middle shaft.The shaft was slightly bent; this condition could help to the fail of the device.No post operative x-rays were provide; therefore the embedded allegation cannot be confirmed.The broken fragment was not returned for evaluation.However, the fracture surface was examined, and no damage related to material issues was observed.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.Properly handling and attention to the approved use of the device diminishes the risk of failure.A dimensional inspection was unable to be performed due to the post-manufacturing damage.A functional test was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the conf intro ndle, dia, 11g 4" would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review a manufacturing record evaluation was performed for the finished device product code: 283903411 lot number: 421175 it was electronically reviewed and no nonconformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: 20 jan 2025 manufacturing site: jabil le locle expiry date: 31 dec 2029 if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONF INTRO NDLE, DIA, 11G 4"
Common Device Name
NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle
SZ  
Manufacturer Contact
elba bello
325 paramount drive
raynham, MA 02767
9083863534
MDR Report Key21890139
Report Number1526439-2025-00511
Device Sequence Number15490464
Product Code GAA
UDI-Device Identifier10705034209456
UDI-Public(01)10705034209456(17)291231(10)421175
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number283903411
Device Lot Number421175
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/28/2025
Supplement Date Received by Manufacturer05/16/2025
Initial Report FDA Received Date04/23/2025
Supplement Report FDA Received Date05/19/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/20/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-