ARTHREX, INC. ANGLED REAMER, ORIENTATION HANDLE; MANUAL INSTR, GENERAL SURGICAL Back to Search Results
Model Number ANGLED REAMER, ORIENTATION HANDLE
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/28/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On 03/28/2025, it was reported by a sales representative via (b)(6) that an ar-9597-20 reamer sleeve fused into the ar-9678 orientation handle and would not spin.This occurred during use in a case with no patient effect.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6.Complaint allegation is not confirmed.One unpackaged ar-9678 angled reamer orientation handle batch number: 052317 was received for investigation.Visual evaluation of the returned device noted no apparent issues with the exterior of the device.Functional testing was performed by assembling the returned device with the returned ar-9597-20 angled reamer sleeve and it was found that the ar-9678 securely locked and unlocked as intended with the ar-9597-20.No problem found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGLED REAMER, ORIENTATION HANDLE
Common Device Name
MANUAL INSTR, GENERAL SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key21890355
Report Number1220246-2025-01661
Device Sequence Number3434529
Product Code LXH
UDI-Device Identifier00888867324947
UDI-Public00888867324947
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 08/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberANGLED REAMER, ORIENTATION HANDLE
Device Catalogue NumberAR-9678
Was Device Available for Evaluation? Yes
Distributor Facility Aware Date03/28/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/28/2025
Supplement Date Received by Manufacturer08/13/2025
Initial Report FDA Received Date04/23/2025
Supplement Report FDA Received Date08/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-