BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE S; IMPLANTABLE LEAD Back to Search Results
Model Number 0137
Medical Device Problem Codes Signal Artifact/Noise (1036); Failure to Capture (1081); Failure to Read Input Signal (1581); Data Problem (3196)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/05/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that right ventricular (rv) lead was exhibiting intermittent loss of capture (loc).Frequent premature ventricular contraction (pvc) caused fusion beats, and an asystole was inappropriately seen by the device.The lead remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE S
Common Device Name
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key21891364
Report Number2124215-2025-25730
Device Sequence Number12798589
Product Code LWS
UDI-Device Identifier00802526531101
UDI-Public(01)00802526531101(17)210913
Combination Product (Y/N)Y
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P910073
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/13/2021
Device Model Number0137
Device Catalogue Number0137
Device Lot Number300967
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/05/2025
Initial Report FDA Received Date04/23/2025
Date Device Manufactured09/13/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age67 YR
Patient SexMale
-
-