• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ESSENTIO SR; IMPLANTABLE PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ESSENTIO SR; IMPLANTABLE PACEMAKER Back to Search Results
Model Number L100
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.A high current drain was detected, but the battery still supported full device function.Detailed analysis confirmed the identified high current drain was a result of compromised capacitors.This resulted in the observed premature battery depletion.It was determined that the capacitors were compromised due to the presence of excess hydrogen in the device case.Boston scientific issued a field safety notice regarding a subset of devices in the accolade pacemaker family that have a potential of exhibiting this behavior.This particular device is included in the hydrogen induced premature depletion advisory population.
 
Event or Problem Description
This product was returned to the manufacturer without any report of performance issues or adverse patient effects.The returned device was analyzed and initial investigation noted a failed longevity calculation.Detailed analysis confirmed premature battery depletion due to high hydrogen induced leaky capacitors.
 
Additional Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.A high current drain was detected, but the battery still supported full device function.Detailed analysis confirmed the identified high current drain was a result of compromised capacitors.This resulted in the observed premature battery depletion.It was determined that the capacitors were compromised due to the presence of excess hydrogen in the device case.Boston scientific issued a field safety notice regarding a subset of devices in the accolade pacemaker family that have a potential of exhibiting this behavior.This particular device is included in the hydrogen induced premature depletion advisory population.This supplemental report was created to capture correction in h6 impact codes.
 
Event or Problem Description
This product was returned to the manufacturer without any report of performance issues or adverse patient effects.The returned device was analyzed and initial investigation noted a failed longevity calculation.Detailed analysis confirmed premature battery depletion due to high hydrogen induced leaky capacitors.Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.A high current drain was detected, but the battery still supported full device function.Detailed analysis confirmed the identified high current drain was a result of compromised capacitors.This resulted in the observed premature battery depletion.It was determined that the capacitors were compromised due to the presence of excess hydrogen in the device case.Boston scientific issued a field safety notice regarding a subset of devices in the accolade pacemaker family that have a potential of exhibiting this behavior.This particular device is included in the hydrogen induced premature depletion advisory population.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ESSENTIO SR
Common Device Name
IMPLANTABLE PACEMAKER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key21891558
Report Number2124215-2025-25756
Device Sequence Number8097755
Product Code LWP
UDI-Device Identifier00802526558900
UDI-Public(01)00802526558900(17)170324
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
N970003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Remedial Action Recall,Notification
Type of Report Initial,Followup
Report Date (Section B) 05/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/24/2017
Device Model NumberL100
Device Catalogue NumberL100
Device Lot Number701469
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/26/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2025
Supplement Date Received by Manufacturer05/23/2025
Initial Report FDA Received Date04/23/2025
Supplement Report FDA Received Date05/23/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/24/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0367-2018
Patient Sequence Number1
Patient Age84 YR
Patient SexFemale
-
-