This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11 additional narrative: h4: the device manufacture date is currently unavailable.Investigation summary = > the complaint device was received at the service center and evaluated.During the service evaluation the following defects were identified: upon receipt : device damaged housing per service reports, this complaint was confirmed.Based on service manual operational and diagnostic, the device was found to be non-repairable, and it was placed into long term hold.Based on the investigation findings, the potential root cause is traced to faulty parts.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|
|
It was reported that the vapr vue wireless footswitch device was returned without an alleged allegation.During in-house engineering evaluation it was determined that there were broken inserts on the housing.It was noted that the inserts crack down the middle, become stripped, or occasionally broken off entirely.There were no adverse consequences to the patient.No additional information could be provided.
|