SMITH & NEPHEW, INC. TWINFIX 2.8 TI W/NDL 22-0 ULTRABRAID; STAPLE, FIXATION, BONE Back to Search Results
Catalog Number 72202067
Medical Device Problem Codes Break (1069); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/31/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: internal complaint reference number: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Event or Problem Description
It was reported that during an unspecified arthroscopy, the twinfix 2.8 ultrabraid anchor got broken.The procedure was completed with a smith and nephew back up device.There was a surgical delay greater than 30 minutes and no further complications were reported.
 
Additional Manufacturer Narrative
H11: internal complaint reference: (b)(4).H2: additional information: h6: medical device problem code.H3, h6: the reported device was received for evaluation.A visual inspection revealed, it was returned in original packaging with the batch number of the complaint on the label.The sutures are still on the device, run through the cannula, and tied off inside the handle.The nosecone has not been rotated to release the sutures, anchor, and needles.The anchor has been fragmented.The proximal end of it is inside the cannula of the insertion device.The distal section is deformed and returned off the insertion device.An osteoraptor anchor and sutures were returned with the twinfix.The osteoraptor and its sutures show no defect.Based on the condition of the product material found during visual inspection, additional material testing is not required.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the material specifications found that a statement (c of c) or evidence of conformance to material and processing specifications must be provided.The root cause has been associated with unintended use of the device.Factors that could have contributed to the failure include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.Based on this investigation, the need for corrective action is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TWINFIX 2.8 TI W/NDL 22-0 ULTRABRAID
Common Device Name
STAPLE, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key21893014
Report Number1219602-2025-00946
Device Sequence Number18239466
Product Code JDR
UDI-Device Identifier03596010619112
UDI-Public3596010619112
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K972326
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72202067
Device Lot Number2091489
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/01/2025
Supplement Date Received by Manufacturer05/22/2025
Initial Report FDA Received Date04/23/2025
Supplement Report FDA Received Date05/23/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/16/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age53 YR
Patient SexFemale
Patient Weight65 KG
-
-