Catalog Number B12XT
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/26/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date sent: 4/23/2025.D4: batch # unk.The device upon which this medwatch is based has been received, however, the evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during an unknown surgery, after insert into the trocar, noted the device was broken.Another device was used to complete the surgery.There was no patient consequence reported.No additional information could be provided.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 5/1/2025.D4: batch # c9e53p.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device visual analysis of the returned sample determined that the b12xt device was returned with no apparent damage.The universal seal was received unlatched from the sleeve.No damage was noted in the sleeve.Upon evaluation of the device had the universal seal latch onto the sleeve assembly.In addition the obturator latch onto the universal seal as well.The event described could not be confirmed as the device was returned without any damaged.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOPATH XCEL BLADELESS TROCAR
Common Device Name
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
3035526892
MDR Report Key21893441
Report Number3005075853-2025-02924
Device Sequence Number15501392
Product Code GCJ
UDI-Device Identifier10705036001195
UDI-Public(01)10705036001195(17)290430(10)116D91
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K032676
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberB12XT
Device Lot Number116D91
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/23/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/26/2025
Supplement Date Received by Manufacturer04/29/2025
Initial Report FDA Received Date04/23/2025
Supplement Report FDA Received Date05/01/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/17/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-